LC KOLOR

Product range / Temperature and humidity data loggers

Temperature and humidity data loggers

Electronic data loggers record complete temperature and humidity profiles during pharmaceutical and food transport, in dry ice down to −100 °C, in warehouses, and during pasteurisation and sterilisation up to +130 °C. Models compliant with EN 12830 and FDA 21 CFR Part 11, with NIST calibration.

Practical applications

Complete temperature and humidity records — from dry ice to pasteurisation

The logger records temperature or humidity throughout the process and generates a report that can be included in quality documentation, claims or audits. We select the model to suit the measurement range, data retrieval method and operating conditions.

Pharmaceuticals and healthcare

GDP-compliant transport of medicines and vaccines, blood banks, clinical trials, pharmacies and wholesalers.

Dry ice and laboratories

Dry ice shipments, −80 °C ultra-low-temperature freezers, biological samples and plasma.

Food industry (HACCP)

Pasteurisation in breweries and bottling plants, canned food, smokehouses, cooking, cold stores and food transport.

Warehouses and indoor spaces

Temperature and humidity monitoring in warehouses, industrial halls, archives and museums.

Sterilisation

Cycle monitoring for autoclaves, sterilisers and washers in healthcare and laboratories.

Temperature data logger in a refrigerated pharmaceutical shipment
Example application — the choice of solution depends on the product, threshold and process conditions.

Full product range details

Temperature & Humidity Data Loggers | LC KOLOR

An electronic temperature data logger records the complete profile of monitored conditions, with a date and time for each measurement. This shows not only whether an excursion occurred, but also when and for how long. These records form the basis of documentation for pharmaceutical transport under Good Distribution Practice (GDP), HACCP systems, thermal process validation and warehouses requiring humidity monitoring.

The range covers four groups: USB transport loggers with automatic PDF reports (iMiniPlus PDF), dry ice loggers down to −100 °C (iMiniPlus Dry Ice), temperature and humidity loggers with alarms (iMINI, iLog), and steel process loggers for pasteurisation, cooking and autoclaves up to +130 °C (KeyTag Kt1HT, Escort Junior). Models meet the requirements of EN 12830 and FDA 21 CFR Part 11, and calibration can be verified with a NIST certificate.

Monitoring vaccine temperatures in a clinic, pharmacy or vaccination centre requires continuous recording within the +2…+8 °C range — including overnight and on non-working days. A logger with an alarm shows when and for how long the refrigerator temperature was outside the required range, while a printout or PDF file provides evidence for public health inspections. The same requirement to document temperature applies to cold rooms in funeral homes, mortuaries and post-mortem facilities.

In breweries and food processing plants, immersion-resistant loggers (IP68) are placed inside a bottle, can or jar passing through a pasteurisation tunnel. Recording the product temperature allows pasteurisation units (PU) to be calculated and process effectiveness to be verified.

Process and pasteurisation data loggers (stainless steel)

2 products
KeyTag Kt1HT — SUS 304 steel high-temperature data logger with two probes

KeyTag Kt1HT

SUS 304 steel high-temperature data logger with two probes

How it works

Highly accurate, reusable data logger made from food-contact-approved stainless steel, with two external probes: a 60 mm top probe (rounded Ø 5 mm tip) and a 25 mm side probe. Used to monitor pasteurisation, cooking, smoking and sterilisation — for example, inside a jar, can or product in an autoclave. When connected via USB, it works like a USB flash drive and automatically generates PDF, CSV and TXT reports. Each unit is calibrated at 4 points (60, 80, 100 and 120 °C) and supplied with a calibration certificate.

Application and use

For measuring temperature inside the product during pasteurisation, cooking, smoking and autoclaving — up to +130 °C. Start the logger before transport and position it as close as possible to the goods being monitored. After delivery, retrieve the data using the selected reporting method.

  • Range −40 °C to +130 °C, accuracy ±0.3 °C across the full range
  • Resolution 0.01 °C, STS40 digital sensor
  • SUS 304 steel housing, IP67 protection
  • PDF report without installing software (magnetic USB connector)
Technical specifications — KeyTag Kt1HT(13)
Technical specification KeyTag Kt1HT
Measurement range−40 °C to +130 °C
Accuracy±0.3 °C (full range)
Resolution0.01 °C
Memoryover 45 500 readings
Logging interval5 s – 24 h
Alarms4 thresholds (continuous and/or cumulative)
Sensor response time20 s (T90, oil bath)
Startvia software, when a temperature is reached or at a set time
FilesPDF, CSV, TXT, KLG — generated automatically
Battery3 V BR2450A, replaceable
Dimensions / weight91.7 × 12.7 × 51 mm / 250 g
SoftwareKeyTag Manager (Windows, macOS, Linux), free
Calibration4 points, certificate included in the PDF report

Data based on the manufacturer's documentation. Full datasheets are available in the documentation below.

Product FAQ

Frequently asked questions

What is KeyTag Kt1HT best suited to?

For measuring temperature inside the product during pasteurisation, cooking, smoking and autoclaving — up to +130 °C.

How do I select the right KeyTag Kt1HT?

The key factors are measurement range, transit time, logging interval, report retrieval method, and the need for alarms and location tracking. Key features of this solution include range −40 °c to +130 °c, accuracy ±0.3 °c across the full range and resolution 0.01 °c, sts40 digital sensor.

Escort Junior — rejestrator pasteryzacji — IP68 waterproof data logger in 316 steel with interchangeable probes

Escort Junior — rejestrator pasteryzacji

IP68 waterproof data logger in 316 steel with interchangeable probes

How it works

A family of high-temperature data loggers up to +125 °C, with a steel base and one or two probes of any length. Probes are easy to unscrew and replace — axial probes of 25, 60 or 80 mm, M6 threaded probes for screwing into a can or jar, and L- and U-shaped probes. Withstands immersion (pressure-tested to 40 m, 5 bar), making it suitable for pasteurisation tunnels, kettles and autoclaves.

Application and use

For pasteurisation of bottles, cans and jars, and submerged operation (IP68) up to +125 °C. Start the logger before transport and position it as close as possible to the goods being monitored. After delivery, retrieve the data using the selected reporting method.

  • Range 0 °C to +125 °C
  • IP68 protection — tested to 40 m / 5 bar
  • 1 or 2 probes, custom lengths available
  • Start with a magnet or at a programmed time
Technical specifications — Escort Junior — rejestrator pasteryzacji(11)
Technical specification Escort Junior — rejestrator pasteryzacji
Measurement range0 °C to +125 °C
Accuracy±0.35 °C (0–70 °C), ±0.40 °C (70–100 °C), ±0.55 °C (100–125 °C)
Resolution0.5 °C
Memory16 000 or 40 000 readings
Logging interval5–59 s or 1–255 min
Housing316 stainless steel and PEEK polymer
Dimensions / weightØ 49 × 28 mm (excluding probe) / 250 g
Batteryhigh-temperature lithium, replaceable, 1–2 years
SoftwareConsole Pro, USB interface
VersionsHJ-IN-V-16-CI (25 mm probe), HJ-FP-V-16-CI / CEI (M6, 60 / 80 mm), HJ-FP-V-16-CU (U-shaped probe), CAP versions
CalibrationNIST

Data based on the manufacturer's documentation. Full datasheets are available in the documentation below.

Manufacturer documentation (original, EN)(1)

Original manufacturer documentation (in English).

Product FAQ

Frequently asked questions

What is Escort Junior — rejestrator pasteryzacji best suited to?

For pasteurisation of bottles, cans and jars, and submerged operation (IP68) up to +125 °C.

How do I select the right Escort Junior — rejestrator pasteryzacji?

The key factors are measurement range, transit time, logging interval, report retrieval method, and the need for alarms and location tracking. Key features of this solution include range 0 °c to +125 °c and ip68 protection — tested to 40 m / 5 bar.

USB transport data loggers with PDF reports

1 product
iMiniPlus PDF — USB data logger with automatic PDF reports — no software required

iMiniPlus PDF

USB data logger with automatic PDF reports — no software required

How it works

Data logger for refrigerated and pharmaceutical shipments, particularly those sent to recipients without software. When connected by USB cable, the computer recognises it as a USB flash drive. The PDF report, containing a graph, a list of readings and an alarm summary, opens in a standard PDF viewer. The large display shows the current, minimum and maximum temperatures.

Application and use

For refrigerated and pharmaceutical shipments where the recipient needs to open a PDF report without installing specialist software. Start the logger before transport and position it as close as possible to the goods being monitored. After delivery, retrieve the data using the selected reporting method.

  • Range −40 °C to +70 °C, resolution 0.1 °C
  • PDF report without installing software
  • HIGH HIGH, HIGH, LOW and LOW LOW alarms
  • Compliant with EN 12830 and FDA 21 CFR Part 11
Technical specifications — iMiniPlus PDF(12)
Technical specification iMiniPlus PDF
Measurement range−40 °C to +70 °C
Accuracy±0.5 °C (−40 to −10 °C), ±0.3 °C (−10 to +70 °C)
Resolution0.1 °C
Sensorinternal or external
Logging interval5 s – 18 h
Clock±1 min per month
Start / stopSTART button (with or without delay), stop function can be disabled
Ingress protectionIP65 (with USB cap fitted)
Dimensions / weight83 × 57 × 17 mm / 70 g
VersionsMU-IN-D-16-L (internal), MU-OE-D-16-L (external), MU-ST-D-8-L (single-use)
Warranty24 months (reusable versions, excluding battery)
CertificationsEN 12830, 21 CFR Part 11, ISO 9001:2015, CE

Data based on the manufacturer's documentation. Full datasheets are available in the documentation below.

Manufacturer documentation (original, EN)(1)

Original manufacturer documentation (in English).

Product FAQ

Frequently asked questions

What is iMiniPlus PDF best suited to?

For refrigerated and pharmaceutical shipments where the recipient needs to open a PDF report without installing specialist software.

How do I select the right iMiniPlus PDF?

The key factors are measurement range, transit time, logging interval, report retrieval method, and the need for alarms and location tracking. Key features of this solution include range −40 °c to +70 °c, resolution 0.1 °c and pdf report without installing software.

Data loggers for dry ice and deep-freeze applications

1 product
iMiniPlus Dry Ice — Dry ice data logger with a probe rated down to −100 °C

iMiniPlus Dry Ice

Dry ice data logger with a probe rated down to −100 °C

How it works

Data logger for dry ice shipments (−78.5 °C) and ultra-low-temperature freezers. The steel probe on a 1-metre Teflon cable goes inside the container, while the logger and its display remain outside. Used for vaccines, biological material, laboratory samples and plasma.

Application and use

For dry ice shipments and ultra-low-temperature freezers — down to −100 °C. Start the logger before transport and position it as close as possible to the goods being monitored. After delivery, retrieve the data using the selected reporting method.

  • Range −100 °C to +40 °C
  • Stainless steel probe on a 1 m Teflon cable
  • Single-use or reusable version
  • Compliant with EN 12830 and FDA 21 CFR Part 11
Technical specifications — iMiniPlus Dry Ice(12)
Technical specification iMiniPlus Dry Ice
Measurement range−100 °C to +40 °C
Accuracy±1.0 °C (−100 to −80 °C), ±0.7 °C (−79 to −10 °C), ±0.6 °C (−9 to +40 °C)
Resolution0.1 °C
Memory8 048 readings
Logging interval5 s – 18 h
Alarms2 thresholds: HIGH and LOW
Probe response timeT90 — 5 min
Ingress protectionIP64 (with cap fitted)
Dimensions / weight83 × 57 × 17 mm / 70 g
Data retrievalvia USB–serial cable, Console Pro software
VersionsMP-OE-N-8-L (reusable), MP-ST-N-8-L (single-use)
CertificationsEN 12830, 21 CFR Part 11, ISO 9001:2015, CE, RoHS

Data based on the manufacturer's documentation. Full datasheets are available in the documentation below.

Manufacturer documentation (original, EN)(1)

Original manufacturer documentation (in English).

Product FAQ

Frequently asked questions

What is iMiniPlus Dry Ice best suited to?

For dry ice shipments and ultra-low-temperature freezers — down to −100 °C.

How do I select the right iMiniPlus Dry Ice?

The key factors are measurement range, transit time, logging interval, report retrieval method, and the need for alarms and location tracking. Key features of this solution include range −100 °c to +40 °c and stainless steel probe on a 1 m teflon cable.

Temperature and humidity (RH) data loggers

2 products
iMINI — temperature and humidity — Data logger with 4 alarm thresholds and LED indicators

iMINI — temperature and humidity

Data logger with 4 alarm thresholds and LED indicators

How it works

Versatile data logger for warehouses, cold stores, transport, blood banks and clinical trials. HS/HE versions measure temperature and humidity simultaneously, while an additional external probe allows temperature measurement inside the product. Two red and two blue LEDs immediately indicate excursions.

Application and use

For warehouses, cold stores and transport requiring temperature and humidity measurement with four alarm thresholds. Start the logger before transport and position it as close as possible to the goods being monitored. After delivery, retrieve the data using the selected reporting method.

  • Temperature −40 °C to +80 °C, humidity 0–100% RH
  • 4 alarm thresholds: 2 red and 2 blue LEDs
  • Up to 3 sensors: internal and external
  • Single-use and reusable versions
Technical specifications — iMINI — temperature and humidity(12)
Technical specification iMINI — temperature and humidity
Temperature range−40 °C to +80 °C
Humidity range0–100% RH
Temperature accuracy±0.5 °C (−40 to −10 °C), ±0.3 °C (−10 to +80 °C)
Humidity accuracy±3% (10–90% RH), ±4% (below 10% and above 90%)
Resolution0.1 °C / 0.1%
Logging interval5 s – 17 h
Start delay1 minute – 99 days
Ingress protectionIP65 (with USB cap fitted)
Dimensions / weight83 × 57 × 17 mm / 70 g
SoftwareConsole Plus; PDF, TXT, CSV export
VersionsMX-IN, MX-ST, MX-OE, MX-1E, MX-2E (temperature), MX-HS, MX-HE (temperature and humidity)
Certifications21 CFR Part 11, ISO 9001:2008, CE, RoHS, NIST

Data based on the manufacturer's documentation. Full datasheets are available in the documentation below.

Manufacturer documentation (original, EN)(1)

Original manufacturer documentation (in English).

Product FAQ

Frequently asked questions

What is iMINI — temperature and humidity best suited to?

For warehouses, cold stores and transport requiring temperature and humidity measurement with four alarm thresholds.

How do I select the right iMINI — temperature and humidity?

The key factors are measurement range, transit time, logging interval, report retrieval method, and the need for alarms and location tracking. Key features of this solution include temperature −40 °c to +80 °c, humidity 0–100% rh and 4 alarm thresholds: 2 red and 2 blue leds.

iLog — temperature and humidity — Data logger with audible alarm, 4 ranges and accessories

iLog — temperature and humidity

Data logger with audible alarm, 4 ranges and accessories

How it works

Round data logger with a large display and visual and audible alarms. Interchangeable external probes cover ranges from dry ice temperatures to +150 °C, while the HS version also measures humidity. Accessories available for iLog include a wall mount, magnetic key, alarm module with siren and strobe, and the portable iLog Chart Reader chart printer — allowing reports to be printed without a computer.

Application and use

For continuous monitoring of rooms and cold stores, with an audible alarm, external siren and report printing without a computer. Start the logger before transport and position it as close as possible to the goods being monitored. After delivery, retrieve the data using the selected reporting method.

  • Ranges: C −20…+100 °C, D −40…+70 °C, N −100…+40 °C, Z 0…+150 °C and RH
  • Memory for 32 000 readings, accuracy ±0.35 °C
  • Visual and audible alarms
  • Accessories: siren with strobe, Chart Reader printer
Technical specifications — iLog — temperature and humidity(12)
Technical specification iLog — temperature and humidity
Temperature rangesC: −20…+100 °C, D: −40…+70 °C, N: −100…+40 °C, Z: 0…+150 °C
HumidityEI-HS-D-32-L version
Accuracy±0.35 °C
Resolution0.1 °C / 0.1%
Memory32 064 readings (1 sensor), 16 032 per sensor (2 sensors)
Logging interval5 s – 10 days
Startwith magnetic key or at a set time
Ingress protectionIP67 (internal sensor), IP65 (external sensor)
Dimensions / weightØ 76 × 18 mm / 68 g
InterfaceUSB (EA-INT-USB) or serial (EA-INT-U), Console Pro software
AccessoriesEA-KEY, EA-ALK, EA-ALARM, EA-INT-HLD, EI-CR
CertificationsEN 12830, 21 CFR Part 11, ISO 9001:2008, CE, RoHS, NIST

Data based on the manufacturer's documentation. Full datasheets are available in the documentation below.

Manufacturer documentation (original, EN)(1)

Original manufacturer documentation (in English).

Product FAQ

Frequently asked questions

What is iLog — temperature and humidity best suited to?

For continuous monitoring of rooms and cold stores, with an audible alarm, external siren and report printing without a computer.

How do I select the right iLog — temperature and humidity?

The key factors are measurement range, transit time, logging interval, report retrieval method, and the need for alarms and location tracking. Key features of this solution include ranges: c −20…+100 °c, d −40…+70 °c, n −100…+40 °c, z 0…+150 °c and rh and memory for 32 000 readings, accuracy ±0.35 °c.

Questions to consider before choosing

Frequently asked questions

What is the difference between a temperature data logger and a temperature indicator?

An indicator only shows whether a threshold has been exceeded. A data logger records each measurement with a date and time, producing a graph and report covering the entire shipment or process.

Which logger generates a PDF report without installing software?

When connected via USB, iMiniPlus PDF and KeyTag Kt1HT work like USB flash drives and automatically generate PDF reports. The other models are read using the free Console Plus or Console Pro software.

Which data logger should I choose for dry ice?

iMiniPlus Dry Ice measures from −100 °C to +40 °C. The probe on a Teflon cable goes inside the container, while the logger and its display remain outside. The iLog N-range model is also suitable for dry ice.

Do the loggers comply with EN 12830 and FDA 21 CFR Part 11?

Yes — iMiniPlus PDF, iMiniPlus Dry Ice and iLog have declared compliance with EN 12830 and 21 CFR Part 11, and iMINI with 21 CFR Part 11. Calibration can be verified with a NIST certificate.

How can I measure pasteurisation temperature inside a bottle or can?

Escort Junior has M6 threaded probes that screw into the packaging, while KeyTag Kt1HT has a 60 mm probe for insertion into the product. Both loggers have steel housings and are sealed (IP68 / IP67).

Free quotation

Enquire about: Temperature and humidity data loggers

We help you choose the right variant and threshold for your application — we usually prepare a quotation within 1 working day. You can also call: +48 22 666 82 95.

Send enquiry

Other product groups